Size of Each Unit (W*D*H)(mm)3U(19"),483*404*130
-
AC InputAC 220V ±10% / 50Hz
-
Input Active Power400W
-
ResolutionAD:24bit;DA:16bit
-
Input Impedance≥1GΩ
-
Voltage
Constant Voltage Range Control: 10mV~5V;
Minimum Discharge Voltage: Upper and lower clamps can discharge to 1V, and 2m wire length can reach 1.5V;
Accuracy: ± 0.01% of FS;
Stability: ± 0.02% of FS;
-
Current
Output Range per Channel: 0.2mA~100mA;100mA~500mA;500mA~3A ;3A~6A;
Accuracy: ± 0.1% of FS;
Constant Voltage Cut-off Current: 0.1mA;0.5mA;3mA; 6mA;
Stability: ± 0.02% of FS;
-
Power
Single Channel Maximum Output Power: 30W;
Stability: ± 0.02% of FS;
-
Time
Current Response Time (10%FS~90%FS)
<1.5ms;
Step Time Range
① ≤(365*24)h/step;② Time Format Support 00:00:00:00(h, min, s, ms)
-
Data Record
Data Recording Conditions
Minimum Time Interval: 10ms;
Minimum Voltage Interval: 5mV;
Minimum Current Interval: 0.1mA;0.5mA;3mA;6mA
Data Acquisition Frequency
100Hz;
-
Charge
Charging Mode: Constant Current Charging, Constant Voltage Charging, Constant Current/ Constant Voltage Charging, Constant Power Charging;
Cut-off Conditions: Voltage, Current, Relative Time, Capacity, -△V;
-
Discharge
Discharge Mode: Constant Current Discharge, Constant Voltage Discharge, Constant Current/ Voltage Discharge, Constant Power Discharge and Constant Resistance Discharge;
Cut-off Conditions: Voltage, Current, Relative Time, and Capacity;
-
Pulse Mode
Charging: Constant Current Mode, Constant Power Mode;
Discharge: Constant Current Mode, Constant Power Mode;
Minimum Pulse Width: 500ms;
Pulse Number: a single pulse step supports 32 different pulses;
Continuous Switching between Charging and Discharging: One Pulse Step can Realize Continuous Switching from Charging to Discharging;
Cut-off Conditions: Voltage, Relative time;
-
DCIR TestSupport DCIR calculating with user defined point.
-
Cycle Life
Cycle Test Range:1~65535 Times;
Single Cycle Steps:254;
Loop nesting: it has the function of nested loop and supports up to 3 levels of nesting;
-
Protect
Power down Data Protection;
Function of Offline Test;
The safety protection conditions can be set, and the setting specs include: upper voltage limit, lower voltage limit, upper current limit, lower current limit, upper capacity limit and delay time.
-
IP Protection DegreeProtection Degree: IP20
-
Channel CharacteristicsConstant current and constant voltage adopt double closed loop structure.
-
Channel Control ModeIndependent Control
-
Voltage and Current SamplingFour Wire Connection
-
Noise<55dB
-
DatabaseAdopting MySQL database to centralized manage testing data.
-
Communication Mode of Upper ComputerBased on TCP/ IP Protocol
-
Server Operating SystemWindows 7 or Windows 10
-
Data Output ModeEXCEL2003,2010、TXT
-
Server Disk Configuration500GB
-
Communication Interface网口
-
Leakage Current<5μA
-
Total Number of Channels8
-
Working Temperature Range0℃~40℃(Within the range of 25±10℃, accuracy drift is 0.005% of FS /℃)
-
Storage Temperature Range-10℃~50℃
-
Relative Humidity Range of Working Environment≤70% RH (There is no condensation of moisture.)
-
Relative Humidity Range of Storage Environment≤80% RH (There is no condensation of moisture.)
-
Clamp TypePolymer Clamps
-
Clamp Picture