Volume, size and weight
-
Nominal Internal Volume25L
-
Inside DimensionsW360 mm×D300 mm×H235 mm
-
Overall DimensionW440 mm×D580 mm×H410 mm
-
WeightAbout 45kg
-
Performance
-
Testing environment
Ambient temperature: +25 ℃, relative humidity: ≤ 85%, there is no sample in the test chamber (no load)
-
Temperature Range15~60℃
-
Temperature Fluctuation≤1℃ (Equivalent to ± 0.5 ℃, no load with stable temperature)
-
Temperature Deviation±2.0℃ (No Load and Stable Temperature)
-
Heating Time25℃→60℃ ≤50 min (No load, average nonlinearity)
-
Cooling Time25℃→15℃ ≤60 min (No load, average nonlinearity)
-
Structure Characteristics
-
Thermal insulation enclosure
Outer wall material: high quality cold rolled steel plate, surface sprayed with plastic and painted.
Inner wall material: stainless steel plate SUS304.
Chamber insulation material: polyurethane foam.
-
Air Conditioning ChannelDC fan, semiconductor refrigeration (heating) module
-
Standard Configuration of Test Chamber
Chamber door: hollow tempered glass and frame;
Cell clamp connecting plate: 2 pieces;
Lighting: LED lights.
-
Control PanelTouch control button
-
Air conditioning unit
Semiconductor refrigeration (heating) module
-
Electric Control System
-
ControllerLED digital display and touch key controller.
-
Setting MethodTouch key type
-
Control ModeForced circulation ventilation. The control system controls the output of the semiconductor refrigeration (heating) module through the PID automatic calculation output results according to the set temperature value, so as to achieve dynamic balance.
-
Communication ModeEthernet Standard Interface
-
Cell Test Equipment and Test Interconnection
-
Testing EquipmentUp to 2 units totaling 16CH (mA Equipment)
-
Middle MachineOne
-
Network SwitchOne
-
Using Conditions: The following conditions are guaranteed by the user (the user is responsible for the installation of the power supply lines of the equipment)
-
Installed Fields
Well ventilated;
No strong vibration around the equipment;
No strong electromagnetic field around the equipment;
No flammable, explosive, corrosive materials and dust around the equipment.
-
Environment Condition
Temperature: 25℃±3℃;
Relative humidity: < 85%;
Pressure: 86kPa~106 kPa.
-
Power Supply/ Capacity/ Maximum Current
AC(220±10%)V /50Hz or AC(110±10%)V /60Hz (Single Phase and Protected Earth Wire)
0.2kW
1A (220V) or 2A (110V)
-
OthersOpening the door of the test chamber while testing will cause temperature fluctuation in the chamber.
-
Cell Specification and Placement Method
-
Cell SpecificationCoin Cell (milliampere level)
-
Cell Placement Mode
Plug on the left and right walls of the inner chamber.